Laboratory Accreditation with NAAU: Step-by-Step Procedure, Application Documents, Deadlines and Common Assessor Findings

NAAU laboratory accreditation step by step: the 17-document file, deadlines for each stage, the UAH 5,400 man-day rate and surveillance. Check your readiness.

Published November 17, 202620 min read
Denys Honchar

By: Denys Honchar

CEO, Ekontrol

NAAU laboratory accreditation: a quality manager reviewing the application file in a testing laboratory

Laboratory Accreditation with NAAU: the Legal Basis Behind the Procedure

Laboratory accreditation in Ukraine means the National Accreditation Agency of Ukraine (NAAU) formally confirms that your testing or calibration laboratory is competent to perform specific methods within a defined scope. The certificate is valid for five years, and NAAU's documents spell out the route to it right down to the number of days you get to sign a contract. Below we walk through that route step by step: what to submit, how long each stage is allowed to take, and where laboratories most often trip up.

There's one core statute: Law of Ukraine No. 2407-III "On Accreditation of Conformity Assessment Bodies" of 17 May 2001, last amended by Law No. 4831-IX of 7 April 2026. Four of its articles matter for a laboratory:

  • Article 10 lists testing and calibration laboratories among the bodies NAAU accredits;
  • Article 11 sets out eight accreditation stages, fixes a five-year validity for the accreditation certificate and leaves NAAU to set its own work deadlines;
  • Article 15 lets NAAU set its fees, but only within a methodology approved by the economy ministry;
  • Article 9 gives you the right to challenge a decision before the appeals commission or in court.

NAAU's own status and functions are set by the Regulation on NAAU, approved by the economy ministry (Article 6 of the law). From there, the agency's own documents take over. Laboratories are assessed against a single standard: DSTU EN ISO/IEC 17025:2019, identical to ISO/IEC 17025:2017 (we cover the standard clause by clause in our complete ISO 17025 guide). The procedure itself sits in Section 8, "Accreditation and Monitoring Procedure", of the NAAU Management Manual, currently revision 62.02 of 14 July 2026. Every deadline in this article comes from that document, with extra detail from the NAAU policies we cite further down.

Why should a business care? NAAU is an associate member of the European co-operation for Accreditation (EA) and, through EA BLA agreements, is recognised for testing and calibration laboratories among other schemes. Since 7 November 2025 it has also been a member of the Global Accreditation Cooperation (Global ACI), which took over the functions of ILAC and IAF on 1 January 2026. That's why NAAU no longer signs new agreements for the combined ILAC MRA mark from 2026.

Competition, by the way, is real. According to the summary table of the NAAU register of accredited bodies, 750 testing laboratories and 40 calibration laboratories held valid certificates as of 23 October 2026.

Initial Accreditation, Re-accreditation or Scope Extension: What Exactly Are You Applying For

Before you start assembling folders for laboratory accreditation, pin down the type of application. NAAU distinguishes three scenarios, and each comes with its own cover letter to the main "Application for Accreditation of a CAB" form (F-08.00.02).

  • Initial accreditation. The laboratory is coming to NAAU for the first time or returning after its certificate was revoked. It means the fullest document set and the longest rework windows.
  • Re-accreditation. The application goes in no later than nine months before the certificate expires (clause 8.7 of the Manual). Miss that date and the assessment follows the rules for initial accreditation.
  • Scope extension. New methods, test items or parameters. NAAU runs an unscheduled assessment, but only if every planned surveillance visit has already taken place.

There's another rule people tend to forget. Want accreditation under more than one scheme, say as both a testing and a calibration laboratory? Each scheme needs its own application. Even the binders differ: NAAU asks for a testing laboratory's file in a black lever-arch binder and a calibration laboratory's in an orange one.

And one unpleasant detail. If NAAU refuses accreditation, it won't accept a new application until a year after the decision. The same applies to laboratories whose certificate was revoked after a surveillance or extraordinary assessment. A year of downtime for a laboratory serving a food plant or a grain elevator costs more than any preparation would. You can see how we work with such laboratories on our Laboratories page.

Laboratory Accreditation Stages: What Happens Between the Application and the Certificate

The law lists eight stages; the NAAU Manual breaks them into concrete actions with deadlines. Here's how it looks from the laboratory's side.

1. The laboratory accreditation application and the review contract

You submit the application, the document set from the official list and the Integrity Commitment (F-08.00.88). NAAU registers the application and sends a draft contract for the application review, which you have 30 calendar days to sign. Note the wording in the application form itself: the laboratory undertakes to pay for the accreditation work regardless of the outcome.

2. Completeness check

An incomplete or incorrectly filled-in file goes back for rework: up to 90 calendar days for initial accreditation and extensions, up to 30 days for re-accreditation. Miss the window and NAAU stops the work, so you'd have to apply again from scratch. At this stage the agency also checks its own resources: if it has no expert in the relevant technical field, it may decline to carry out the accreditation.

3. Assessment team, contracts, programme

NAAU appoints an accreditation assessment team. For a first-time applicant the team has at least two people. You're told who's on it in advance, and you can object to its composition under policy ZD-08.00.15. The accreditation work contracts must be signed within 30 days, and NAAU draws up the accreditation programme within 10 days.

4. Document review

The team reviews your file off-site and sends you a document review report. The Manual sets no deadline for the review itself. If there are findings, you get up to 90 days to revise for initial accreditation or up to 60 days for re-accreditation. There can be no more than two repeat reviews, each under an additional agreement. In practice, this is where most of the time gets lost.

5. Assessment plan and on-site assessment

NAAU sends the on-site assessment plan, and the laboratory has three days to accept or reject it. Then comes the assessment itself: an opening meeting, evidence gathering, analysis and a closing meeting where nonconformities and correction deadlines are recorded.

6. Corrective actions

You get up to 90 calendar days to clear nonconformities after an initial accreditation assessment and up to 60 days after a re-accreditation one. If NAAU finds the evidence insufficient, you have just 10 days to submit more. There can be no more than two repeat on-site assessments.

7. Decision and accreditation certificate

The NAAU director takes the decision once the on-site assessment report and the full file are in. The grounds for refusal are stated plainly: late or ineffective correction of nonconformities, more than one serious violation from list ZD-08.00.35, or evidence of fraud or concealed information. If everything's in order, you receive the accreditation certificate with its "Scope of Accreditation" annex and sign the Accreditation Agreement (F-08.00.27). The certificate is issued in Ukrainian; an English version with the same registration number is available on written request.

So how long does it all take? Adding up the maximum deadlines gives 73 calendar days with no rework and 353 days with every rework loop, and that excludes document review, the on-site assessment and the decision, which have no deadlines at all. NAAU's own FAQ (published in 2016) points to 6 to 12 months. Our experience matches that: the laboratories that finish in under six months are the ones that submitted a clean file the first time.

The maximum deadlines for initial accreditation add up to 73 calendar days with no rework and 353 days with every rework loop; document review, the on-site assessment and the decision have no set deadline (Source: NAAU Management Manual extract, rev. 62.02, clause 8.2).
StageInitial accreditation and extensionRe-accreditationManual clause
Signing the application review contractup to 30 daysup to 30 days8.2.1
Completing an incomplete application fileup to 90 daysup to 30 days8.2.2
Signing the accreditation work contractsup to 30 daysup to 30 days8.2.3
Accreditation programme drafted (NAAU)10 days10 days8.2.3
Revising documents after the review reportup to 90 days, no more than 2 repeat reviewsup to 60 days8.2.4
Agreeing the on-site assessment plan3 days3 days8.2.5
Corrective actions after the on-site assessmentup to 90 daysup to 60 days8.2.7
Extra evidence if corrective actions fall short10 days10 days8.2.7
Consequence of a missed deadlineWork stopped, new applicationWork stopped, apply as for initial accreditation8.2.2–8.2.7

Pre-assessment: one day, once only, no consulting

If both sides agree, NAAU can run a pre-assessment to see whether the laboratory is ready for assessment (methodology M-08.00.10). It lasts one day and can be held only once per applicant. Assessors don't consult during it, so don't expect tips on how to rewrite a procedure. For a first accreditation with a broad scope we'd recommend it: one day on site often saves months of repeat document reviews.

Documents for Laboratory Accreditation: NAAU's 17-Item List

The document list for a testing laboratory is set by form F-08.17.03 (revision 07 of 14 July 2026). You can download it with all the templates from NAAU's "Applying for accreditation" page (in Ukrainian). Calibration laboratories have their own set of forms, but the logic is the same.

The submission format matters too. The application, questionnaire, laboratory passport, draft scope, self-assessment checklist and integrity commitment go in both on paper and as Word files on a USB stick. The charter, laboratory regulations, register extracts and org chart are filed as scans plus paper copies. Management system documentation, copies of diplomas and test reports, and method evaluation data are submitted electronically, while the location details and floor plan are paper-only. For re-accreditation, floor plans and diplomas are only needed if staff, status or operations have changed.

It helps to look at the file from the other side: which document proves which clause of the standard. The laboratory passport and the method evaluation data carry most of the technical weight. The self-assessment checklist F-08.17.20 works as a cross-reference index: for each clause from 4 to 8 you enter the number of your own document that meets the requirement.

Measurement uncertainty (clause 7.6), however, has no dedicated form in the file. The assessor will look for it in the checklist and in your management system documentation, so your uncertainty budget needs to be ready when you apply, not when the team arrives. We explain how to build one under the GUM in our article on measurement uncertainty and calibration.

The laboratory passport and method evaluation data cover most technical clauses of the standard, while uncertainty (clause 7.6) has no dedicated form (Source: NAAU forms F-08.17.03, F-08.17.19, F-08.17.20, F-08.17.33; mapping by Ekontrol).
No.Document (NAAU form)What the assessor looks for
1Application for accreditation (F-08.00.02)Type of work, legal entity, sites, staff, independence statement
2Certified copy of the charterThat the lab is a legal entity or a defined part of one (ISO/IEC 17025 clause 5.1)
3Laboratory regulationsReporting lines and the manager's authority
4Certified extract from the state registerCurrent registration details, any bankruptcy proceedings
5Tax status documentDetails for the contract and invoices
6Questionnaire (F-08.17.15)Addresses, subcontractors, other accreditations, volume of work in the year before applying
7Laboratory passport (F-08.17.19), forms 1–10Equipment, reference materials, staff, premises, methods, subcontracting, proficiency testing
8Draft scope of accreditation (F-08.17.17)Methods, test items, parameters and ranges
9Management system documentationThe full set of procedures for clauses 4–8
10Location details of the premisesWhere exactly the on-site assessment will take place
11Floor plan showing equipment layoutWhether the premises suit the types of testing
12Copies of diplomas and qualification recordsEducation and technical competence of each analyst
13Copies of test reports or sampling recordsActual work in each area of activity
14Self-assessment checklist (F-08.17.20)A reference to your own document for every ISO/IEC 17025 clause
15Test method evaluation data (F-08.17.33)Verification or validation, equipment traceability, monitoring of result validity
16Org chart and staffing table without salariesRoles, deputies, adequacy of staffing
17Integrity commitment (F-08.00.88)An NAAU anti-corruption requirement, not a clause of the standard

Scope of Accreditation: How to Draft One That Survives Document Review

Laboratories often fill in the draft scope of accreditation (F-08.17.17) last, even though it drives the size and cost of the whole job. NAAU has a dedicated instruction, "Drafting the scope of accreditation of a testing laboratory" (IN-08.01.03), and it's worth reading before you enter the first method.

Three rules that get checked on site:

  1. Every method is backed by actual work. NAAU's criteria for sustained experience (ZD-08.00.19) require the laboratory to have performed tests under every method in the requested scope within the last two years. For initial accreditation, the on-site assessment must cover all methods, and if there isn't a single case file for an area of activity, that area will be refused.
  2. Every area has interlaboratory comparison results. NAAU's proficiency testing policy (ZD-08.00.29, rev. 05) requires satisfactory proficiency testing (PT) or other interlaboratory comparison (ILC) results for each area of activity in the draft scope before accreditation is granted. NAAU points to the EPTIS database for finding schemes. Where a PT provider is accredited by NAAU and its scope overlaps with yours, taking part in its programmes is mandatory.
  3. Every method has evaluation data. Form F-08.17.33 asks, for each method, whether it's standardised or in-house, who performs it, on what equipment and with what proof of traceability, how result validity is monitored, and for a reference to the verification report (standardised methods) or validation report (non-standard and modified methods).

We give a short overview of the standard's own requirements for method selection, verification and validation (clause 7.2) on our ISO/IEC 17025 page.

Our advice is simple, if not very welcome: a narrower scope you can prove in full beats a wide one with loose ends. A refused area can only come back through a scope extension, and NAAU runs those only after the planned surveillance visits.

A method with no test reports in two years is a candidate for removal

A familiar story: a method goes into the scope "just in case" because the lab once bought equipment for it. During the document review or the on-site visit the team asks for test reports from the last two years, and there aren't any. The outcome is predictable: the scope gets restricted under the sustained-experience criterion. If the business really needs the method, run real tests with it, issue reports and take part in PT before you apply.

On-site Assessment: How NAAU Assessors Work and How to Prepare Your Staff

The on-site assessment is the main stage of laboratory accreditation (for laboratories it follows methodology M-08.01.14). The team has already seen your documents; now it checks whether the system actually runs the way it's written.

What happens in practice:

  • interviews with the manager, quality manager and analysts, with checklists filled in as they go;
  • witnessing one of your analysts perform a test under a method from the scope;
  • sampling records from the customer request through to the test report, checking that every result can be traced;
  • inspecting premises, equipment, environmental conditions and the records kept on them;
  • reviewing PT and other interlaboratory comparison results.

A few formalities. The team is always photographed at the opening and closing meetings, while photos or video of equipment and work are only taken with your permission. Observers from the laboratory side need to be agreed with NAAU in advance. If the laboratory has branches, initial accreditation covers both the head office and every branch.

A word on martial law. The Manual allows remote assessments for laboratories in frontline regions and combined ones, where part of the team is on site and part works online. The NAAU director decides on such an assessment based on a reasoned internal memo (methodology M-08.00.23). Don't count on a remote assessment by default.

Prepare your staff for a conversation, not a procedures exam. The assessor is unlikely to ask for a clause number. They'll ask you to show how you check whether a reagent is fit for use, or what you do when a control chart goes out of limits. A good rehearsal is an internal audit under ISO 17025 run as thoroughly as NAAU would run it.

Common NAAU Assessor Findings: What We See in Practice

A caveat first: we haven't found published nonconformity statistics in NAAU's documents, so this section draws on the Ekontrol team's experience preparing and supporting laboratories.

  1. The scope is wider than the actual work. Methods with no reports in two years: the equipment's there, the orders aren't.
  2. PT as a box-ticking exercise. One round for the entire scope, no participation plan for the cycle and no analysis of z-scores, even though ZD-08.00.29 explicitly requires those records.
  3. Verification of a standardised method with no data. There's an order to "implement the standard", but no evidence the laboratory achieves the method's performance characteristics (repeatability, limit of detection, trueness). Modified methods often lack validation (ISO/IEC 17025 clause 7.2).
  4. Uncertainty estimated but never used. The budget sits in a folder, yet it's nowhere to be seen in test reports or in the decision rules for statements of conformity (clause 7.8.6).
  5. Traceability via verification certificates. Where a method needs the uncertainty of the reference standard, a legal-metrology verification certificate won't provide it. We explain the difference in Calibration vs Verification.
  6. Impartiality in a lab owned by a manufacturer. Impartiality risks are written up in a single line, while the production manager whose output gets rejected sits one floor up (clause 4.1.4).
  7. Authorisation without evidence. An analyst performs a method, but there are no records of training, competence checks or authorisation for that specific method (clause 6.2.5).
  8. The passport doesn't match the scope. An instrument the method needs isn't listed in form 3, or a room from form 6 lacks the environmental control the method requires.

Most of these findings show up where the system was written for the audit and the day-to-day work was forgotten. An experienced team spots that within half a day.

How Much Does Laboratory Accreditation with NAAU Cost

There's no fixed price list for laboratory accreditation, and that isn't the agency being difficult. The cost is calculated under the Methodology for Determining the Cost of Accreditation and Monitoring Work, approved by Ministry of Economic Development Order No. 973 of 5 September 2012, using a simple formula:

cost = time standard in man-days × cost of one man-day.

NAAU sets the man-day rate by its own order. The current rate, in force since 5 June 2026, is UAH 5,400 (NAAU Order No. 18od of 3 June 2026). The time standards for each stage (application review, document review, on-site assessment, decision, issuing the certificate) are set in a separate fee procedure (NAAU Order No. 8od of 11 February 2022, as amended) and adjusted by scope complexity coefficients. That's why two laboratories with the same headcount can receive very different invoices: what counts is the number of methods and areas in the scope, the number of sites and the size of the assessment team.

What else to budget for:

  • the methodology sets costs excluding VAT, which is charged separately under the law;
  • travel and subcontracted services aren't included in the rate and appear as a separate line;
  • a repeat document review is formalised in an additional agreement and has its own time standard in NAAU's tables;
  • surveillance visits and the reassessment over the five years are paid too, since Article 12 of the law obliges accredited bodies to pay for monitoring;
  • payment doesn't depend on the outcome, a commitment you make in the application itself.

NAAU states the exact amount in the draft contract after reviewing the application. Our preparation services are billed separately from NAAU's fees. And honestly, laboratories lose the most money not on consultants or on NAAU, but on repeat reviews and repeat assessments caused by an unfinished file.

After the Certificate: Surveillance, Scope Extension and Re-accreditation Over the Five-Year Cycle

The accreditation certificate is valid for five years, and NAAU calls that period the accreditation cycle. Throughout the cycle the agency monitors the laboratory through surveillance visits, unscheduled and extraordinary assessments, and a reassessment. The surveillance schedule is set by the NAAU Monitoring Policy (ZD-08.00.16, revision 12); the main first-cycle dates are summarised in the table below.

A few cycle rules worth putting in the laboratory's internal calendar:

  • nonconformities from a surveillance visit must be cleared within 30 calendar days of the end of the on-site assessment, otherwise NAAU suspends the certificate in full or for part of the scope;
  • if the laboratory doesn't sign the surveillance contract on time or doesn't pay, the certificate is suspended and only reinstated after an extraordinary on-site assessment;
  • at the start of every January the laboratory reports to NAAU on the previous year: volume of work by type of test, staff, reports issued and complaints;
  • a suspension lasts no more than a year, after which the certificate is revoked, and a revoked certificate can't be reinstated.

Laboratories that have operated for five years without suspensions for serious violations get evened-out 15-month intervals in later cycles, so the first surveillance moves from month 12 to month 15.

Scope extension

An extension uses the same F-08.00.02 application form, with a cover letter for extension and a separate document list (for testing laboratories, F-08.17.40). NAAU carries out an unscheduled assessment under methodology M-08.00.18, and the deadlines for revisions and corrective actions are the same as for initial accreditation: up to 90 days.

Re-accreditation

Treat re-accreditation as a project in its own right. The application and document set go in no later than nine months before the certificate expires, and the reassessment must take place no later than three months before that date. The rework windows are shorter here: 30 days for the application file, 60 for the documentation and 60 for corrective actions. Miss them and NAAU stops the work, leaving the laboratory to apply as if for initial accreditation.

Cycle eventWhen it happens (first cycle)Source document
Start of the cycleDate of the accreditation decision or a later certificate start dateNAAU Manual, clause 8.2.8
First surveillance12 months after the certificate start dateZD-08.00.16, item 2
Second surveillance16 months after the planned date of the first (around month 28)ZD-08.00.16, item 2
Third surveillanceAnother 16 months later (around month 44)ZD-08.00.16, item 2
Allowed deviation from the scheduleOne month earlier or later than plannedZD-08.00.16, item 7
Re-accreditation applicationNo later than 9 months before expiry (month 51)NAAU Manual, clause 8.7
ReassessmentNo later than 3 months before expiry (month 57)ZD-08.00.16, item 9
Certificate expiryMonth 60, i.e. after 5 yearsLaw No. 2407-III, Art. 11

Laboratory Accreditation Readiness Checklist Before You Apply

Run through this list before submitting your laboratory accreditation application. If you answer "no" to three or more items, it's wiser to hold the application back for a month or two: rework on NAAU's deadlines will cost you more.

  1. The application type is clear, there's a separate application for each scheme, and for re-accreditation more than 9 months remain on the certificate.
  2. All 17 documents from list F-08.17.03 are ready, using the current templates from the NAAU website.
  3. Every method in the draft scope has test reports from the last two years.
  4. Every area has satisfactory PT or ILC results and a participation plan for the accreditation cycle.
  5. Form F-08.17.33 is completed for each method, with a reference to the verification or validation report.
  6. Uncertainty is estimated for every quantitative method and shows up in test reports and decision rules.
  7. Equipment listed in the passport is calibrated with metrological traceability, and calibration certificates have been reviewed.
  8. Impartiality risks are identified specifically, with actions.
  9. Every analyst has records of training, competence checks and authorisation for specific methods.
  10. A full internal audit against clauses 4–8 and a management review have been completed.
  11. Checklist F-08.17.20 points to real documents, not to "Quality Manual, Section 4".
  12. The budget covers NAAU fees for five years: accreditation, three surveillance visits and re-accreditation.

If you'd like an independent view, start with an independent readiness check of your laboratory. It shows how many items on this list are genuinely closed and how many exist only on paper.

After that, we can provide support during the NAAU assessment, from answering the document review report to corrective actions after the visit. That way one person owns the NAAU deadlines while the laboratory keeps working as usual.

What actually shortens the road to the certificate

The biggest time saving comes from a clean file on the first submission, because each round of document revisions can cost up to 90 days. A scope where every method is backed by reports and PT matters just as much. For a first accreditation, add the NAAU pre-assessment. Polished manuals and thick binders hardly move the timeline at all.