What ISO 17025 is and what it's called in Ukraine
ISO 17025 (full designation ISO/IEC 17025:2017) is the international standard that sets general requirements for the competence, impartiality and consistent operation of testing and calibration laboratories. Laboratory managers usually come to us with three questions: do we have to, how much will it cost, and how long will it take. The standard only answers part of that: it describes what a laboratory must look like, while the law and NAAU decide whether it's mandatory, how long it takes and what it costs, so this guide covers laboratory accreditation from all three angles.
The standard applies equally to an independent contract lab and to the in-house lab of a dairy plant or a grain elevator, whatever the headcount. A laboratory whose ISO 17025 competence has been confirmed by NAAU (the National Accreditation Agency of Ukraine) receives an accreditation certificate, and its reports are trusted without retesting. How we prepare labs for that step is described on our ISO 17025 accreditation support page.
The international text was published in November 2017: third edition, 30 pages, developed by ISO/CASCO. According to the standard's page on iso.org, it was last reviewed and confirmed in 2023, so the 2017 edition remains current.
DSTU EN ISO/IEC 17025:2019: the Ukrainian version
Ukraine applies an identical national standard, DSTU EN ISO/IEC 17025:2019 (EN ISO/IEC 17025:2017, IDT; ISO/IEC 17025:2017, IDT). The national standards body UkrNDNC adopted it by Order No. 483 of 23 December 2019; it has been in force since 1 January 2021 and replaced DSTU ISO/IEC 17025:2017. Searches for "dstu 17025", "dstu iso/iec 17025" or the Cyrillic "ісо 17025" all point to the same document: the content matches the international text, only the designation and language differ. NAAU's accreditation procedure (Management Manual, revision 62.02 of 14 July 2026) names DSTU EN ISO/IEC 17025:2019 as the criterion for testing and calibration laboratories.
How ISO 17025 differs from ISO 9001
ISO 9001 checks whether an organisation is under control. ISO/IEC 17025 asks a harder question: can anyone trust your number? An ISO 9001 certificate doesn't prove that a lab evaluates uncertainty correctly or that its balances are traceable to SI, so it doesn't replace accreditation, although a working QMS gives you a head start in clause 8 (Option B).
We cover ISO 9001 for the whole company in our complete ISO 9001 guide.
There's no such thing as an “ISO 17025 certificate”
Laboratories aren't certified to ISO 17025; they're accredited. The difference is fundamental: under Law No. 2407-III only the national accreditation body, NAAU, issues accreditation certificates. For official food control, an "ISO 17025 certificate of conformity" from a private company means nothing, because Law No. 2042-VIII recognises only laboratories accredited by NAAU or by a foreign accreditation body that meets the law's criteria. You can check any laboratory in the NAAU register by its certificate number.
Laboratory accreditation: when it's mandatory and when it's voluntary
Let's start with something consultants don't always say out loud: there's no general obligation for a Ukrainian laboratory to get accredited. Law No. 2407-III "On Accreditation of Conformity Assessment Bodies" lists voluntary accreditation among its principles (Art. 5) while also covering cases where accreditation is "required by legislation" (Art. 2). The law itself defines who accredits (Art. 6: NAAU only) and whom (Art. 10, where testing and calibration laboratories come first). The obligation always comes from another act or from a contract.
Here's where it shows up in practice:
- Official control of food and feed. Law No. 2042-VIII defines an accredited laboratory as one accredited to ISO/IEC 17025 by NAAU, by a foreign body that's a full ILAC member, or by a body meeting ISO/IEC 17011 (Art. 1(1)). An authorised laboratory doing tests for official control is, by the same law's definition, an accredited one (Art. 1(46)).
- Calibration. Law No. 1314-VII "On Metrology and Metrological Activity" allows measuring instruments to be calibrated by scientific metrology centres, metrology centres and NAAU-accredited laboratories, as well as by labs with documented traceability of their standards (Art. 27(2)). Accreditation is one of two routes here.
- Legal verification you can skip. Article 17 of the same law exempts legally controlled instruments from periodic verification when an accredited lab uses them within its scope of accreditation. For a lab with dozens of balances and thermometers, that's a real saving.
- Conformity assessment under technical regulations. For designating bodies in this area NAAU keeps a separate general document, "Provisions on accreditation for designation purposes in the field of technical regulations" (ZD-08.03.37). If your reports are meant to support such an assessment, check the requirements of the specific regulation.
- Contracts. The most common case, and the law doesn't touch it at all: an export contract, a tender or a retail chain accepts reports only from an accredited laboratory.
| Situation | Is ISO/IEC 17025 accreditation required? | Basis |
|---|---|---|
| Authorised laboratory for official control of food and feed | Yes, authorisation isn't possible without it | Law No. 2042-VIII, Art. 1(1) and 1(46) |
| Calibration laboratory calibrating instruments for others | No, the alternative is documented traceability of standards | Law No. 1314-VII, Art. 27(2) |
| Exempting your own legally controlled instruments from periodic verification | Yes, within the scope only | Law No. 1314-VII, Art. 17(1) |
| In-house production lab for internal control | No, unless a customer or certification scheme requires it | Voluntary principle, Law No. 2407-III, Art. 5 |
| Reports for export, tenders, retail chains | Depends on the contract | Contract, customer requirements |
| Medical laboratory | A different standard applies: ISO 15189:2022 | NAAU Management Manual, cl. 8.1.3 |
ISO 17025 in Ukraine: how many laboratories hold NAAU accreditation
NAAU publishes its register of accredited conformity assessment bodies as separate files for each scheme. We counted entries with an unexpired certificate and no suspension or withdrawal mark. The result: 749 testing and 40 calibration laboratories accredited to ISO/IEC 17025 (files dated 23 October and 15 September 2026), plus 123 medical laboratories accredited to ISO 15189. Together with certification, inspection and verification bodies that's 1,233 valid entries, and ISO 17025 laboratories make up almost two thirds of them.
The more interesting number is elsewhere. The testing laboratory file has 1,484 rows, and almost half of them (735) relate to certificates that have been withdrawn, suspended or simply expired. Getting accredited is realistic. Keeping it for five years is harder, and it's worth thinking about before you apply.
If your lab isn't in the register yet, sector-specific preparation notes are on our laboratories page.
DSTU 17025 requirements: what's assessed in clauses 4–8
Clauses 1–3 are introductory (scope, normative references, terms); the requirements start in clause 4. We counted the lowest-level numbered clauses: there are 141 of them, spread very unevenly.
- Clause 4. General requirements (9 clauses). Impartiality and confidentiality. The lab has to identify risks to its impartiality on an ongoing basis, including those arising from ownership or reporting lines (4.1.4). For a plant lab testing its own factory's products, this is the most sensitive point.
- Clause 5. Structural requirements (7 clauses). A legal entity or a defined part of one, the range of laboratory activities, management responsibilities.
- Clause 6. Resource requirements (31 clauses). Personnel, facilities and environmental conditions, equipment (13 clauses on its own), metrological traceability, externally provided products and services.
- Clause 7. Process requirements (69 clauses). Review of requests and contracts, selection, verification and validation of methods, sampling, technical records, uncertainty, ensuring the validity of results, reporting (19 clauses on reports and certificates alone), complaints, nonconforming work, data management.
- Clause 8, the management system (25 clauses), gives you a choice. Option A: the lab meets clauses 8.2–8.9 (documents, records, risks, improvement, corrective action, internal audits, management review). Option B: if the lab runs an ISO 9001 system and can consistently meet clauses 4–7, it also meets 8.2–8.9 (8.1.3).
The planning takeaway is simple: half of the requirements sit in clause 7. You can write a quality policy in a week, but not a method validation or an uncertainty budget, so build your schedule around the technical work.
| ISO/IEC 17025 clause | What's assessed | Evidence you'll be asked for |
|---|---|---|
| 4. General | Risks to impartiality, confidentiality commitments | Impartiality risk analysis, signed staff commitments |
| 5. Structure | Legal status, range of activities, allocation of responsibilities | Articles or lab regulations, org chart, job descriptions |
| 6. Resources | Staff competence and authorisation, environmental conditions, calibration | Competence matrix, authorisation records, calibration schedule, certificate reviews |
| 7. Process | Methods, uncertainty, quality control, reports, decision rule | Verification and validation reports, uncertainty budgets, control charts, proficiency testing results |
| 8. Management system | Documents and records, risks, corrective actions, internal audits, management review | Internal audit programme and reports, corrective action log, management review minutes |
Personnel, equipment and traceability under ISO 17025
Clause 6 is often underrated because it looks like housekeeping. In reality it's where assessors check that every result is backed by a competent person and a fit-for-purpose instrument.
Personnel: competence plus authorisation
The standard requires procedures and records for determining competence requirements, selection, training, supervision, authorisation and monitoring of staff competence (6.2.5). Separately, the lab authorises people for specific activities: developing, modifying, verifying and validating methods, analysing results including statements of conformity, and reporting, reviewing and authorising results (6.2.6). A diploma in a personnel file proves nothing here: the assessor wants to see who cleared a technician for a given method, when, and on what basis.
Equipment and calibration
Measuring equipment has to be calibrated when measurement accuracy or uncertainty affects the validity of results, or when calibration is needed to establish metrological traceability (6.4.6). A calibration certificate sitting in a binder doesn't close the requirement. The lab has to review it, compare the error and uncertainty with its own requirements and decide whether the instrument is fit for use. We explain how calibration differs from legal verification, and which of the two the law requires, in our article on calibration vs verification under ISO 17025.
Metrological traceability
Traceability to SI is established through a documented, unbroken chain of calibrations (6.5.1, 6.5.2). NAAU's traceability policy (ZD-08.00.09, revision 10) requires labs to decide on the need for calibration through a risk analysis using DSTU ILAC-G 24/OIML D 10, taking the manufacturer's requirements and actual operating conditions into account. In other words, "once a year for everything" isn't a requirement: it's the lab's decision, and it has to be justified.
A certificate without an accreditation reference doesn't prove traceability
Under NAAU's traceability policy, a calibration certificate issued without reference to accreditation by a body covered by the mutual recognition arrangements is treated as a service from a supplier whose competence hasn't been confirmed. That result isn't sufficient evidence of traceability. An exception is possible for calibrations not covered by recognition arrangements; the lab then has to gather the evidence itself, and NAAU may assess it, up to an audit of the non-accredited calibration lab with NAAU representatives present. It's simpler to choose accredited providers from the start and check their scope in the register.
Method validation, uncertainty and proficiency testing
Clause 7 is the biggest, and it's where labs that copied someone else's documentation most often stumble.
Method verification and validation
Before using a method, the lab must confirm it can perform it properly and keep verification records (7.2.1.5). Validation is required for non-standard methods, laboratory-developed methods, and standard methods used outside their intended scope or otherwise modified (7.2.2.1). In practice labs mix the two up in both directions: they either validate a standard method across forty pages or "verify" their own modification with a single parallel measurement.
Measurement uncertainty
The lab identifies the contributions to uncertainty and accounts for all significant ones, including those arising from sampling (7.6.1). A calibration lab evaluates uncertainty for all calibrations, including its own equipment (7.6.2); a testing lab evaluates it too, and where the method doesn't allow a rigorous calculation, estimates it from the method's principles or practical experience (7.6.3). We walked through building a GUM and EA-4/02 budget for a real instrument in our article on measurement uncertainty and calibration.
Validity of results and proficiency testing
Clause 7.7.1 lists internal quality control options (reference materials, control charts, replicate testing), and 7.7.2 requires comparison with other labs through proficiency testing (PT) or interlaboratory comparisons where available and appropriate. NAAU makes this concrete in its policy ZD-08.00.29 (revision 05): satisfactory PT results covering the applicant's scope are needed before accreditation is granted, and afterwards the lab takes part at least once for each area of activity during the accreditation cycle. Skipping an area without justification can cost you part of your scope.
Decision rule
When a report includes a statement of conformity to a specification, the lab documents the decision rule it applies, taking into account the risk of false acceptance or false rejection (7.8.6.1). Agree it with the customer at contract stage; otherwise a result close to the tolerance limit turns into a dispute after the report is issued.
How NAAU accreditation works: stages and deadlines
Law No. 2407-III (Art. 11) lists the stages: application, documents, appointment of the assessment team, work programme, document review, on-site assessment, an assessment report with recommendations, and the decision. The law leaves timelines to NAAU, which sets them in its accreditation and monitoring procedure (section 8 of the Management Manual, revision 62.02 of 14 July 2026, published on the For our clients page).
NAAU posts the application forms on its accreditation application page. For a testing laboratory the package includes the application (F-08.00.02), an integrity commitment, a list of documents, a questionnaire, the draft scope of accreditation, the laboratory passport, a checklist and a method evaluation data form. The scope takes the longest: every object, parameter and method has to be described in a way an assessor can verify.
A few rules from the Manual worth knowing in advance:
- a preliminary assessment is possible by mutual agreement, lasts one day, happens once and involves no consulting;
- for a first-time applicant, the assessment team has at least two members;
- initial accreditation covers the head office and every branch;
- during hostilities, remote and combined assessments are allowed for labs in regions where staff safety is at risk;
- after a refusal, a new application can only be filed a year later.
We set out the detailed step-by-step procedure, with documents and typical assessor findings, in our guide to NAAU laboratory accreditation step by step.
| Stage | Deadline (NAAU documents) | What it means for the lab |
|---|---|---|
| Contract for application review | Up to 30 calendar days to review and sign | Agree the budget before you apply so internal sign-offs don't cost you a month |
| Completing the application package | Up to 90 days (initial), up to 30 days (reaccreditation) | Missing the deadline means a decision to stop the work and a new application |
| Accreditation programme | 10 calendar days | Covers every clause of the standard, the full scope and all sites |
| Rework after document review | Up to 90 days (initial), up to 60 days (reaccreditation), no more than two re-reviews | The Manual doesn't provide for a third attempt |
| Corrective actions after the on-site assessment | Up to 90 days (initial), up to 60 days (reaccreditation); extra evidence within 10 days | Plan resources for corrective actions right after the closing meeting |
| Accreditation certificate | 5 years (accreditation cycle) | First cycle: surveillance after 12 months, then twice at 16-month intervals |
| Reaccreditation | Documents no later than 9 months before the certificate expires | If you're late, the procedure runs as an initial one |
Preparing your lab for ISO 17025 accreditation?
We'll run a gap analysis against clauses 4–8, help you define the scope of accreditation, validate methods, build uncertainty budgets and a proficiency testing plan, and then support you through the NAAU assessment.
Discuss your lab's preparationHow much laboratory accreditation costs and how long it takes
The official part of the cost is calculated transparently, even though only NAAU can give you the final figure in advance. The law requires an accredited body to pay for all accreditation and monitoring work (Art. 12), while NAAU sets the fees under a methodology approved by the ministry and makes them public (Art. 15).
The formula: the number of person-days under NAAU's workload norms (Procedure and Amount of Payment for Accreditation and Monitoring Work, NAAU Order No. 8od of 11 February 2022, as amended), multiplied by the cost of a person-day. Since 5 June 2026 a person-day costs UAH 5,400 (NAAU Order No. 18od of 3 June 2026, published in the NAAU orders section). Expenses tied to specific work, such as assessor travel, come on top. The number of person-days depends on your scope: methods, areas, sites. So NAAU calculates the exact amount from your application, and we'd be sceptical of any "turnkey accreditation price" quoted without a scope.
Besides NAAU fees, budget for your own costs, which are easy to underestimate: calibration at accredited labs, PT participation, reference materials, training and specialist time for validation. These depend on the scope, so we deliberately don't quote averages.
Timelines
The law says plainly that NAAU sets accreditation timelines (Art. 11). The deadlines in the table above mostly apply to the applicant: up to 90 days to complete the documents and up to another 90 days for corrective actions after the assessment. Those two windows alone add up to half a year a lab can burn on its own if it applies before it's ready. The real timeline depends less on the agency than on the state of your system on the day you apply.
The most honest way to estimate it is a preliminary readiness assessment, once you can see which validations, uncertainty budgets and PT results are still missing.
Common nonconformities in ISO 17025 assessments
NAAU doesn't publish nonconformity statistics by clause, so what follows is our experience preparing laboratories, not a ranking:
- Impartiality as a box-ticking exercise. There's a risk register, but it misses the main risk for a plant lab: production pressure on the results of testing its own products.
- No authorisation records. A technician runs a method, but there's no document clearing that person for that method (6.2.6).
- Calibration certificates are filed, but nobody compared the error and uncertainty with the method's tolerance.
- Traceability relies on a non-accredited provider with no evidence of its competence.
- A standard method wasn't verified before use (7.2.1.5), or the verification doesn't support the performance claimed in the scope.
- Uncertainty budgets don't cover the scope. One parameter was calculated, twenty were declared.
- The PT plan doesn't cover every area of the scope, although NAAU requires participation in each area during the cycle.
- The decision rule wasn't agreed with the customer, yet the report states "conforms".
- Spreadsheets and the LIMS weren't checked for functionality before use (7.11.2), and access isn't restricted.
- The internal audit has no technical part: the auditor reviewed documents but didn't witness a single test.
Many of these overlap with management system nonconformities in general, which we collected in our top 15 ISO audit nonconformances.
The last one is fixed through training. We covered the technical side of internal audits, including witnessing a test, at our seminar on DSTU EN ISO/IEC 17025:2019 and internal laboratory audits.
You can lose your certificate between assessments
NAAU documents spell out the grounds for suspending a certificate: findings from a surveillance visit not closed within 30 calendar days, or a lab that doesn't sign monitoring contracts on time or doesn't pay for that work (in which case reinstatement requires an extraordinary on-site assessment). PT skipped without justification puts part of the scope at risk. The protection is simple: a cycle calendar with surveillance dates, a PT plan and a reaccreditation filing date 9 months before the certificate expires.
ISO 17025 for your laboratory: where to start
Once the decision to get accredited is made, our projects usually follow this sequence:
- Scope. A list of objects, parameters and methods your customers actually need. A broad first scope means more person-days, uncertainty budgets and PT.
- Gap analysis against clauses 4–8, with particular attention to clauses 6 and 7.
- Technical groundwork: method verification and validation, uncertainty budgets, a calibration plan with certificate reviews.
- Management system documents and records under Option A, or additions to an existing QMS under Option B.
- A PT plan for the cycle, using NAAU-accredited providers and the EPTIS database that NAAU itself refers to.
- An internal audit and management review: we recommend completing at least one full cycle before applying, so you can show records as well as procedures.
- An application to NAAU with the full set of forms.
To be honest about expectations: ISO 17025 doesn't make a lab more accurate by itself. It makes you prove that your numbers can be reproduced and checked, and that takes months of technical work, not weeks of paperwork.
If you'd rather avoid going round in circles, we run management system development and implementation together with your team, not instead of it.
To scope the work for your specific methods, get in touch: we'll start with your method list and the state of your equipment.
What accreditation gives you besides the certificate
First, international recognition of your reports: NAAU is recognised by EA for testing and calibration laboratories and, since 7 November 2025, has been a full member of the Global Accreditation Cooperation (Global ACI), which took over the functions of ILAC and IAF on 1 January 2026. Second, exemption from periodic verification for legally controlled instruments within your scope (Art. 17 of Law No. 1314-VII). Third, the option to become an authorised laboratory for official food control under Law No. 2042-VIII.

Need a certification consultation?
Free Consultation
On This Page
- What ISO 17025 is and what it's called in Ukraine
- Laboratory accreditation: when it's mandatory and when it's voluntary
- ISO 17025 in Ukraine: how many laboratories hold NAAU accreditation
- DSTU 17025 requirements: what's assessed in clauses 4–8
- Personnel, equipment and traceability under ISO 17025
- Method validation, uncertainty and proficiency testing
- How NAAU accreditation works: stages and deadlines
- How much laboratory accreditation costs and how long it takes
- Common nonconformities in ISO 17025 assessments
- ISO 17025 for your laboratory: where to start
What ISO 17025 is and what it's called in Ukraine
ISO 17025 (full designation ISO/IEC 17025:2017) is the international standard that sets general requirements for the competence, impartiality and consistent operation of testing and calibration laboratories. Laboratory managers usually come to us with three questions: do we have to, how much will it cost, and how long will it take. The standard only answers part of that: it describes what a laboratory must look like, while the law and NAAU decide whether it's mandatory, how long it takes and what it costs, so this guide covers laboratory accreditation from all three angles.
The standard applies equally to an independent contract lab and to the in-house lab of a dairy plant or a grain elevator, whatever the headcount. A laboratory whose ISO 17025 competence has been confirmed by NAAU (the National Accreditation Agency of Ukraine) receives an accreditation certificate, and its reports are trusted without retesting. How we prepare labs for that step is described on our ISO 17025 accreditation support page.
The international text was published in November 2017: third edition, 30 pages, developed by ISO/CASCO. According to the standard's page on iso.org, it was last reviewed and confirmed in 2023, so the 2017 edition remains current.
DSTU EN ISO/IEC 17025:2019: the Ukrainian version
Ukraine applies an identical national standard, DSTU EN ISO/IEC 17025:2019 (EN ISO/IEC 17025:2017, IDT; ISO/IEC 17025:2017, IDT). The national standards body UkrNDNC adopted it by Order No. 483 of 23 December 2019; it has been in force since 1 January 2021 and replaced DSTU ISO/IEC 17025:2017. Searches for "dstu 17025", "dstu iso/iec 17025" or the Cyrillic "ісо 17025" all point to the same document: the content matches the international text, only the designation and language differ. NAAU's accreditation procedure (Management Manual, revision 62.02 of 14 July 2026) names DSTU EN ISO/IEC 17025:2019 as the criterion for testing and calibration laboratories.
How ISO 17025 differs from ISO 9001
ISO 9001 checks whether an organisation is under control. ISO/IEC 17025 asks a harder question: can anyone trust your number? An ISO 9001 certificate doesn't prove that a lab evaluates uncertainty correctly or that its balances are traceable to SI, so it doesn't replace accreditation, although a working QMS gives you a head start in clause 8 (Option B).
We cover ISO 9001 for the whole company in our complete ISO 9001 guide.
There's no such thing as an “ISO 17025 certificate”
Laboratories aren't certified to ISO 17025; they're accredited. The difference is fundamental: under Law No. 2407-III only the national accreditation body, NAAU, issues accreditation certificates. For official food control, an "ISO 17025 certificate of conformity" from a private company means nothing, because Law No. 2042-VIII recognises only laboratories accredited by NAAU or by a foreign accreditation body that meets the law's criteria. You can check any laboratory in the NAAU register by its certificate number.
Laboratory accreditation: when it's mandatory and when it's voluntary
Let's start with something consultants don't always say out loud: there's no general obligation for a Ukrainian laboratory to get accredited. Law No. 2407-III "On Accreditation of Conformity Assessment Bodies" lists voluntary accreditation among its principles (Art. 5) while also covering cases where accreditation is "required by legislation" (Art. 2). The law itself defines who accredits (Art. 6: NAAU only) and whom (Art. 10, where testing and calibration laboratories come first). The obligation always comes from another act or from a contract.
Here's where it shows up in practice:
- Official control of food and feed. Law No. 2042-VIII defines an accredited laboratory as one accredited to ISO/IEC 17025 by NAAU, by a foreign body that's a full ILAC member, or by a body meeting ISO/IEC 17011 (Art. 1(1)). An authorised laboratory doing tests for official control is, by the same law's definition, an accredited one (Art. 1(46)).
- Calibration. Law No. 1314-VII "On Metrology and Metrological Activity" allows measuring instruments to be calibrated by scientific metrology centres, metrology centres and NAAU-accredited laboratories, as well as by labs with documented traceability of their standards (Art. 27(2)). Accreditation is one of two routes here.
- Legal verification you can skip. Article 17 of the same law exempts legally controlled instruments from periodic verification when an accredited lab uses them within its scope of accreditation. For a lab with dozens of balances and thermometers, that's a real saving.
- Conformity assessment under technical regulations. For designating bodies in this area NAAU keeps a separate general document, "Provisions on accreditation for designation purposes in the field of technical regulations" (ZD-08.03.37). If your reports are meant to support such an assessment, check the requirements of the specific regulation.
- Contracts. The most common case, and the law doesn't touch it at all: an export contract, a tender or a retail chain accepts reports only from an accredited laboratory.
| Situation | Is ISO/IEC 17025 accreditation required? | Basis |
|---|---|---|
| Authorised laboratory for official control of food and feed | Yes, authorisation isn't possible without it | Law No. 2042-VIII, Art. 1(1) and 1(46) |
| Calibration laboratory calibrating instruments for others | No, the alternative is documented traceability of standards | Law No. 1314-VII, Art. 27(2) |
| Exempting your own legally controlled instruments from periodic verification | Yes, within the scope only | Law No. 1314-VII, Art. 17(1) |
| In-house production lab for internal control | No, unless a customer or certification scheme requires it | Voluntary principle, Law No. 2407-III, Art. 5 |
| Reports for export, tenders, retail chains | Depends on the contract | Contract, customer requirements |
| Medical laboratory | A different standard applies: ISO 15189:2022 | NAAU Management Manual, cl. 8.1.3 |
ISO 17025 in Ukraine: how many laboratories hold NAAU accreditation
NAAU publishes its register of accredited conformity assessment bodies as separate files for each scheme. We counted entries with an unexpired certificate and no suspension or withdrawal mark. The result: 749 testing and 40 calibration laboratories accredited to ISO/IEC 17025 (files dated 23 October and 15 September 2026), plus 123 medical laboratories accredited to ISO 15189. Together with certification, inspection and verification bodies that's 1,233 valid entries, and ISO 17025 laboratories make up almost two thirds of them.
The more interesting number is elsewhere. The testing laboratory file has 1,484 rows, and almost half of them (735) relate to certificates that have been withdrawn, suspended or simply expired. Getting accredited is realistic. Keeping it for five years is harder, and it's worth thinking about before you apply.
If your lab isn't in the register yet, sector-specific preparation notes are on our laboratories page.
DSTU 17025 requirements: what's assessed in clauses 4–8
Clauses 1–3 are introductory (scope, normative references, terms); the requirements start in clause 4. We counted the lowest-level numbered clauses: there are 141 of them, spread very unevenly.
- Clause 4. General requirements (9 clauses). Impartiality and confidentiality. The lab has to identify risks to its impartiality on an ongoing basis, including those arising from ownership or reporting lines (4.1.4). For a plant lab testing its own factory's products, this is the most sensitive point.
- Clause 5. Structural requirements (7 clauses). A legal entity or a defined part of one, the range of laboratory activities, management responsibilities.
- Clause 6. Resource requirements (31 clauses). Personnel, facilities and environmental conditions, equipment (13 clauses on its own), metrological traceability, externally provided products and services.
- Clause 7. Process requirements (69 clauses). Review of requests and contracts, selection, verification and validation of methods, sampling, technical records, uncertainty, ensuring the validity of results, reporting (19 clauses on reports and certificates alone), complaints, nonconforming work, data management.
- Clause 8, the management system (25 clauses), gives you a choice. Option A: the lab meets clauses 8.2–8.9 (documents, records, risks, improvement, corrective action, internal audits, management review). Option B: if the lab runs an ISO 9001 system and can consistently meet clauses 4–7, it also meets 8.2–8.9 (8.1.3).
The planning takeaway is simple: half of the requirements sit in clause 7. You can write a quality policy in a week, but not a method validation or an uncertainty budget, so build your schedule around the technical work.
| ISO/IEC 17025 clause | What's assessed | Evidence you'll be asked for |
|---|---|---|
| 4. General | Risks to impartiality, confidentiality commitments | Impartiality risk analysis, signed staff commitments |
| 5. Structure | Legal status, range of activities, allocation of responsibilities | Articles or lab regulations, org chart, job descriptions |
| 6. Resources | Staff competence and authorisation, environmental conditions, calibration | Competence matrix, authorisation records, calibration schedule, certificate reviews |
| 7. Process | Methods, uncertainty, quality control, reports, decision rule | Verification and validation reports, uncertainty budgets, control charts, proficiency testing results |
| 8. Management system | Documents and records, risks, corrective actions, internal audits, management review | Internal audit programme and reports, corrective action log, management review minutes |
Personnel, equipment and traceability under ISO 17025
Clause 6 is often underrated because it looks like housekeeping. In reality it's where assessors check that every result is backed by a competent person and a fit-for-purpose instrument.
Personnel: competence plus authorisation
The standard requires procedures and records for determining competence requirements, selection, training, supervision, authorisation and monitoring of staff competence (6.2.5). Separately, the lab authorises people for specific activities: developing, modifying, verifying and validating methods, analysing results including statements of conformity, and reporting, reviewing and authorising results (6.2.6). A diploma in a personnel file proves nothing here: the assessor wants to see who cleared a technician for a given method, when, and on what basis.
Equipment and calibration
Measuring equipment has to be calibrated when measurement accuracy or uncertainty affects the validity of results, or when calibration is needed to establish metrological traceability (6.4.6). A calibration certificate sitting in a binder doesn't close the requirement. The lab has to review it, compare the error and uncertainty with its own requirements and decide whether the instrument is fit for use. We explain how calibration differs from legal verification, and which of the two the law requires, in our article on calibration vs verification under ISO 17025.
Metrological traceability
Traceability to SI is established through a documented, unbroken chain of calibrations (6.5.1, 6.5.2). NAAU's traceability policy (ZD-08.00.09, revision 10) requires labs to decide on the need for calibration through a risk analysis using DSTU ILAC-G 24/OIML D 10, taking the manufacturer's requirements and actual operating conditions into account. In other words, "once a year for everything" isn't a requirement: it's the lab's decision, and it has to be justified.
A certificate without an accreditation reference doesn't prove traceability
Under NAAU's traceability policy, a calibration certificate issued without reference to accreditation by a body covered by the mutual recognition arrangements is treated as a service from a supplier whose competence hasn't been confirmed. That result isn't sufficient evidence of traceability. An exception is possible for calibrations not covered by recognition arrangements; the lab then has to gather the evidence itself, and NAAU may assess it, up to an audit of the non-accredited calibration lab with NAAU representatives present. It's simpler to choose accredited providers from the start and check their scope in the register.
Method validation, uncertainty and proficiency testing
Clause 7 is the biggest, and it's where labs that copied someone else's documentation most often stumble.
Method verification and validation
Before using a method, the lab must confirm it can perform it properly and keep verification records (7.2.1.5). Validation is required for non-standard methods, laboratory-developed methods, and standard methods used outside their intended scope or otherwise modified (7.2.2.1). In practice labs mix the two up in both directions: they either validate a standard method across forty pages or "verify" their own modification with a single parallel measurement.
Measurement uncertainty
The lab identifies the contributions to uncertainty and accounts for all significant ones, including those arising from sampling (7.6.1). A calibration lab evaluates uncertainty for all calibrations, including its own equipment (7.6.2); a testing lab evaluates it too, and where the method doesn't allow a rigorous calculation, estimates it from the method's principles or practical experience (7.6.3). We walked through building a GUM and EA-4/02 budget for a real instrument in our article on measurement uncertainty and calibration.
Validity of results and proficiency testing
Clause 7.7.1 lists internal quality control options (reference materials, control charts, replicate testing), and 7.7.2 requires comparison with other labs through proficiency testing (PT) or interlaboratory comparisons where available and appropriate. NAAU makes this concrete in its policy ZD-08.00.29 (revision 05): satisfactory PT results covering the applicant's scope are needed before accreditation is granted, and afterwards the lab takes part at least once for each area of activity during the accreditation cycle. Skipping an area without justification can cost you part of your scope.
Decision rule
When a report includes a statement of conformity to a specification, the lab documents the decision rule it applies, taking into account the risk of false acceptance or false rejection (7.8.6.1). Agree it with the customer at contract stage; otherwise a result close to the tolerance limit turns into a dispute after the report is issued.
How NAAU accreditation works: stages and deadlines
Law No. 2407-III (Art. 11) lists the stages: application, documents, appointment of the assessment team, work programme, document review, on-site assessment, an assessment report with recommendations, and the decision. The law leaves timelines to NAAU, which sets them in its accreditation and monitoring procedure (section 8 of the Management Manual, revision 62.02 of 14 July 2026, published on the For our clients page).
NAAU posts the application forms on its accreditation application page. For a testing laboratory the package includes the application (F-08.00.02), an integrity commitment, a list of documents, a questionnaire, the draft scope of accreditation, the laboratory passport, a checklist and a method evaluation data form. The scope takes the longest: every object, parameter and method has to be described in a way an assessor can verify.
A few rules from the Manual worth knowing in advance:
- a preliminary assessment is possible by mutual agreement, lasts one day, happens once and involves no consulting;
- for a first-time applicant, the assessment team has at least two members;
- initial accreditation covers the head office and every branch;
- during hostilities, remote and combined assessments are allowed for labs in regions where staff safety is at risk;
- after a refusal, a new application can only be filed a year later.
We set out the detailed step-by-step procedure, with documents and typical assessor findings, in our guide to NAAU laboratory accreditation step by step.
| Stage | Deadline (NAAU documents) | What it means for the lab |
|---|---|---|
| Contract for application review | Up to 30 calendar days to review and sign | Agree the budget before you apply so internal sign-offs don't cost you a month |
| Completing the application package | Up to 90 days (initial), up to 30 days (reaccreditation) | Missing the deadline means a decision to stop the work and a new application |
| Accreditation programme | 10 calendar days | Covers every clause of the standard, the full scope and all sites |
| Rework after document review | Up to 90 days (initial), up to 60 days (reaccreditation), no more than two re-reviews | The Manual doesn't provide for a third attempt |
| Corrective actions after the on-site assessment | Up to 90 days (initial), up to 60 days (reaccreditation); extra evidence within 10 days | Plan resources for corrective actions right after the closing meeting |
| Accreditation certificate | 5 years (accreditation cycle) | First cycle: surveillance after 12 months, then twice at 16-month intervals |
| Reaccreditation | Documents no later than 9 months before the certificate expires | If you're late, the procedure runs as an initial one |
Preparing your lab for ISO 17025 accreditation?
We'll run a gap analysis against clauses 4–8, help you define the scope of accreditation, validate methods, build uncertainty budgets and a proficiency testing plan, and then support you through the NAAU assessment.
Discuss your lab's preparationHow much laboratory accreditation costs and how long it takes
The official part of the cost is calculated transparently, even though only NAAU can give you the final figure in advance. The law requires an accredited body to pay for all accreditation and monitoring work (Art. 12), while NAAU sets the fees under a methodology approved by the ministry and makes them public (Art. 15).
The formula: the number of person-days under NAAU's workload norms (Procedure and Amount of Payment for Accreditation and Monitoring Work, NAAU Order No. 8od of 11 February 2022, as amended), multiplied by the cost of a person-day. Since 5 June 2026 a person-day costs UAH 5,400 (NAAU Order No. 18od of 3 June 2026, published in the NAAU orders section). Expenses tied to specific work, such as assessor travel, come on top. The number of person-days depends on your scope: methods, areas, sites. So NAAU calculates the exact amount from your application, and we'd be sceptical of any "turnkey accreditation price" quoted without a scope.
Besides NAAU fees, budget for your own costs, which are easy to underestimate: calibration at accredited labs, PT participation, reference materials, training and specialist time for validation. These depend on the scope, so we deliberately don't quote averages.
Timelines
The law says plainly that NAAU sets accreditation timelines (Art. 11). The deadlines in the table above mostly apply to the applicant: up to 90 days to complete the documents and up to another 90 days for corrective actions after the assessment. Those two windows alone add up to half a year a lab can burn on its own if it applies before it's ready. The real timeline depends less on the agency than on the state of your system on the day you apply.
The most honest way to estimate it is a preliminary readiness assessment, once you can see which validations, uncertainty budgets and PT results are still missing.
Common nonconformities in ISO 17025 assessments
NAAU doesn't publish nonconformity statistics by clause, so what follows is our experience preparing laboratories, not a ranking:
- Impartiality as a box-ticking exercise. There's a risk register, but it misses the main risk for a plant lab: production pressure on the results of testing its own products.
- No authorisation records. A technician runs a method, but there's no document clearing that person for that method (6.2.6).
- Calibration certificates are filed, but nobody compared the error and uncertainty with the method's tolerance.
- Traceability relies on a non-accredited provider with no evidence of its competence.
- A standard method wasn't verified before use (7.2.1.5), or the verification doesn't support the performance claimed in the scope.
- Uncertainty budgets don't cover the scope. One parameter was calculated, twenty were declared.
- The PT plan doesn't cover every area of the scope, although NAAU requires participation in each area during the cycle.
- The decision rule wasn't agreed with the customer, yet the report states "conforms".
- Spreadsheets and the LIMS weren't checked for functionality before use (7.11.2), and access isn't restricted.
- The internal audit has no technical part: the auditor reviewed documents but didn't witness a single test.
Many of these overlap with management system nonconformities in general, which we collected in our top 15 ISO audit nonconformances.
The last one is fixed through training. We covered the technical side of internal audits, including witnessing a test, at our seminar on DSTU EN ISO/IEC 17025:2019 and internal laboratory audits.
You can lose your certificate between assessments
NAAU documents spell out the grounds for suspending a certificate: findings from a surveillance visit not closed within 30 calendar days, or a lab that doesn't sign monitoring contracts on time or doesn't pay for that work (in which case reinstatement requires an extraordinary on-site assessment). PT skipped without justification puts part of the scope at risk. The protection is simple: a cycle calendar with surveillance dates, a PT plan and a reaccreditation filing date 9 months before the certificate expires.
ISO 17025 for your laboratory: where to start
Once the decision to get accredited is made, our projects usually follow this sequence:
- Scope. A list of objects, parameters and methods your customers actually need. A broad first scope means more person-days, uncertainty budgets and PT.
- Gap analysis against clauses 4–8, with particular attention to clauses 6 and 7.
- Technical groundwork: method verification and validation, uncertainty budgets, a calibration plan with certificate reviews.
- Management system documents and records under Option A, or additions to an existing QMS under Option B.
- A PT plan for the cycle, using NAAU-accredited providers and the EPTIS database that NAAU itself refers to.
- An internal audit and management review: we recommend completing at least one full cycle before applying, so you can show records as well as procedures.
- An application to NAAU with the full set of forms.
To be honest about expectations: ISO 17025 doesn't make a lab more accurate by itself. It makes you prove that your numbers can be reproduced and checked, and that takes months of technical work, not weeks of paperwork.
If you'd rather avoid going round in circles, we run management system development and implementation together with your team, not instead of it.
To scope the work for your specific methods, get in touch: we'll start with your method list and the state of your equipment.
What accreditation gives you besides the certificate
First, international recognition of your reports: NAAU is recognised by EA for testing and calibration laboratories and, since 7 November 2025, has been a full member of the Global Accreditation Cooperation (Global ACI), which took over the functions of ILAC and IAF on 1 January 2026. Second, exemption from periodic verification for legally controlled instruments within your scope (Art. 17 of Law No. 1314-VII). Third, the option to become an authorised laboratory for official food control under Law No. 2042-VIII.